check quantity

Thermal Conductivity Detector (TCD) with Add-On Kit for PerkinElmerClarus 500 GC, 230V

Please enter valid quantity

Please login to add favorites


Product Number List Price Your Price Quantity
N6520023 3897.00 USD

You successfully added item(s) to your cart


Kit includes all necessary items to install the TCD into theinstrument: detector, heater, sensor, heater block, 1/6 in. gasline, and flow controller pneumatics.
How it Works: Balanced carrier flow is maintained through the analytical andreference channels. The filaments are electrically heated andassume a temperature dependent upon the thermal conductivity of thegas passing over them. When column effluent of a different thermalconductivity passes through the analytical or reference channel,the bridge circuit becomes unbalanced and gives rise to a signalthat yields the chromatographic output.

detectorType TCD
One Unit Contains 1 each
Product Group Add-on Kit