PerkinElmer uses cookies to ensure that we give you the best experience possible on our website. This may include cookies from third party websites. If you continue without changing your settings, we will assume that you consent to receive cookies from this website. You can change your cookie settings at any time. To learn more, please review our cookie policy, which includes information on how to manage your cookies.

Strep-Tactin AlphaLISA Acceptor bead interference

The table below shows results from a compound interference study. We tested a panel of 24 compounds in a QC assay using our AlphaLISA® toolbox bead products. The information below can be used for guidance as to what buffer components might interfere with AlphaLISA assays using Strep-Tactin Acceptor beads, and at what concentrations interference might occur. These data are derived from single experiments and should serve more as a guide rather than a precise value, as it is unlikely you will be using the exact same assay design as used to generate the data below. Interference concentrations may vary depending on the assay components. It is possible your assay will tolerate higher or lower concentration than what is shown.



No effect
10%
50% lossTested up to% of inhibition at max conc
Glycerol%0.211.730%80
DMSO%7.6>1010%38
Gelatin%>1>11%0
CHAPS%0.1>33%46
Deoxycholate%0.0050.33%77
SDS%0.060.13%96
Triton X-100%>3>33%0
Tween 20%>3>33%0
TRIS (pH 8.0)M5.9E-021.3E-030.5M64
2-MercaptoethanolM3.8E-032.9E-020.1M63
DTTM6.6E-061.2E-030.1M80
EDTAM>0.1>0.10.1M0
EGTAM>0.1>0.10.1M0
IBMXM>0.0025>0.0025M0.0025M0
ImidazoleM1.3E-036.8E-030.1M99
Mg2+ (MgCl2)M3.4E-02>0.10.1M44
NitroprussideM2.4E-034.3E-020.1M80
UreaM>0.1>0.10.1M0
VanadateM6.4E-033.8E-020.1M56
CitrateM1.4E-032.6E-030.1M100
Adenosine (ATP)M0.01M>0.01M0.01M0
GlycineM0.01M>0.01M0.01M0
NaFM>0.01M>0.01M0.01M0
S-adenosylmethionineM>0.001>0.0010.001M0
Complete Protease InhibitorXNot determinedNot determined1X0

Refer to the figure below for more detail as to what the values in the table above indicate. Please note that the figure is provided as a general example to explain how the values in our interference tables were determined. The table above indicates that an assay testing the effect of different chemicals using Strep-Tactin AlphaLISA Acceptor beads (with Anti-Rabbit IgG Donor beads) showed little effect on the assay signal at final Triton X-100 concentrations up to 3%. Using 11.7% glycerol resulted in a 50% loss of signal.

interferencetesting16_ASK.jpg