PerkinElmer uses cookies to ensure that we give you the best experience possible on our website. This may include cookies from third party websites. If you continue without changing your settings, we will assume that you consent to receive cookies from this website. You can change your cookie settings at any time. To learn more, please review our cookie policy, which includes information on how to manage your cookies.

Deoxycholate and bead interference

The table below shows results from a compound interference study. We tested a panel of 24 compounds in a QC assay using our AlphaLISA® toolbox bead products. The information below can be used for guidance as to whether deoxycholate might interfere in assays involving particular bead products, and at what deoxycholate concentration interference might occur. These data are derived from single experiments and should serve more as a guide rather than a precise value, as it is unlikely you will be using the exact same assay design as used to generate the data below. Interference concentrations may vary depending on the assay components. It is possible your assay will tolerate higher or lower concentration than what is shown.



Tested up to 3%
Catalog NumberProductNo effect (10%)50% loss% of inhibition at max conc
AL125Streptavidin AlphaLISA Acceptor Beads0.0050.0294
AL126Protein L AlphaLISA Acceptor Beads0.010.0690
AL127Anti-FITC AlphaLISA Acceptor Beads0.010.04100
AL128 (biotin)Anti-6xHis AlphaLISA Acceptor Beads0.0090.0599
AL128 (GSH)Anti-6XHis AlphaLISA Acceptor Beads0.0050.01100
AL129Anti-V5 AlphaLISA Acceptor Beads0.00020.01100
AL130Anti-mouse IgM AlphaLISA Acceptor Beads0.0030.0398
AL131Anti-chicken IgY AlphaLISA Acceptor Beads0.0070.0696
AL132Anti-sheep IgG AlphaLISA Acceptor Beads0.0070.0592
AL133Anti-GFP AlphaLISA Acceptor Beads0.0030.02100
AL134Anti-MBP AlphaLISA Acceptor Beads0.0090.03100
AL136Strep-Tactin AlphaLISA Acceptor Beads0.0050.377
AS102Protein A Alpha Donor Beads0.0040.02100
AS103Anti-FLAG Alpha Donor Beads0.0050.04100
AS104Anti-mouse IgG Alpha Donor Beads0.0040.0297
AS105Anti-rabbit IgG Alpha Donor Beads0.0050.0294
AS106Strep-Tactin Alpha Donor Beads0.030.283
AL900AlphaLISA TruHits kit 2 ug/mL0.020.0487
AL900AlphaLISA TruHits kit 20 ug/mL0.0070.0387

Refer to the figure below for more detail as to what the values in the table above indicate. Please note that the figure is provided as a general example to explain how the values in our interference tables were determined. The table above indicates that an assay testing the effect of deoxycholate using anti-6X His AlphaLISA Acceptor beads (with glutathione Donor beads) showed little effect on the assay signal at final deoxycholate concentrations up to 0.005%. Deoxycholate did show a half-maximal inhibition of signal when used at a final concentration of 0.01%. When 3% deoxycholate was included in the assay, 100% signal inhibition was observed.

interferencetesting16_ASK.jpg
Figure: Key for how the values in our interference tables are derived. The “no effect” concentration was obtained by extrapolating the MAX - 10% MAX counts. Refer to the table above for actual interference values for each toolbox bead product.