Loading . . .

4336

XRpad 4336 Wireless a-Si X-Ray Detector
The PerkinElmer XRpadTM 4336 is a wireless, light weight, cassette-sized flat panel detector for digital radiography.

Overview

The PerkinElmer XRpadTM 4336 is a wireless, light weight, cassette-sized flat panel detector for digital radiography. It fits into a conventional table or wall-stand Bucky, just like a film-screen cassette.

Featuring a 15 million pixel image matrix, a best-in-class 100 μm pixel pitch, and a directly deposited CsI scintillator, the PerkinElmer XRpadTM 4336 provides exceptional image quality. True 35 × 43 cm2 imaging area per the ISO 4090 cassette standard is provided, along with a removable, rechargeable battery. Single-piece carbon-fiber construction for the front and back housing allows for easier placement and cleaning. Automatic Exposure Detection simplifies integration.

We have over 20 years of experience partnering with customers to develop products in a wide range of X-ray applications. Let our digital imaging experience work for you.

Applications
  • Digital Radiography

Features & Benefits

  • Cassette detector per ISO 4090, fits in conventional Bucky
  • True 35 cm × 43 cm (14” × 17”) image
  • High resolution 100 μm pixel pitch
  • Up to 65,536 gray levels
  • Automatic Exposure Detection (AED)
  • Direct deposition Csl, for excellent image quality
  • Wireless 802.11n standard
  • Integrated status display
  • Rechargeable battery
  • Optional connector for power and data communication
  • On board memory for image storage
Disclaimer Icon

PerkinElmer flat panel X-ray detectors are components intended to be integrated into products by X-ray systems manufacturers. Systems manufacturers are responsible for qualifying and validating their products for intended uses and meeting all applicable regulatory requirements.

Specifications

Sensor
PanelSingle substrate amorphous silicon active TFT/diode array
ScintillatorDirect deposition CsI:Tl
Pixel Matrix3556 x 4320
Pixel Pitch100 μm
Electronics
AmplifiersLow noise ASICs with user selectable gains
ADC16 bit
Image Transfer TimeWired: 500 ms; Wireless 3000 ms
On-board Memory1 GB DDR3, 4 GB SDHC card
Mechanical
SizeISO 4090 for 35 cm x 43 cm (14" x 17") cassette size
Active AreaTrue 355 mm x 432 mm
External Dimensions384 mm (w) x 460 mm (l) x 15 mm (h)
Weight3.8 kg (8.4 lbs)
HousingCarbon-fiber front & back
Communications
Status DisplayOLED display with WiFi, battery, and sensor indicators
Wireless Data I/F802.11n WiFi standard
Wired Data I/FGigE via optional power and communication tether
X-ray I/FIntegrated X-ray trigger control
Automatic Exposure Detection
Imaging Performance
Limiting Resolution5 cy/mm
Typical MTF70% (1 cy/mm), 40% (2 cy/mm),15% (4 cy/mm) for RQA5
Typical DQE75% (0 cy/mm), 60% (1 cy/mm), 40% (3 cy/mm) for RQA5
Energy Range20 — 150 kV
Environmental
Temperature10 — 35 °C operating
Humidity30 — 70 % RH operating (non-condensing)
Power
BatteryRechargeable battery, 11.1 V
Battery ChargerExternal 2 bay charger 100 — 240 V AC 50/60 Hz
Interface and Power UnitOptional XRpad IPU with external power supply 100-240 V AC, GigE, and X-ray I/F
Regulatory
StandardsIEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, FCC 47CFR PT 15, FCC OET 65C, ETSI EN 301 893, EN 62311 ISO 10993-5, ISO 10993-10, CE